6-40
IM 701240-01E
6.17 Setting the Wave Window Trigger (ENHANCED)
<For a description of this function, refer to page 2-23.>
Procedure
1. Press MENU.
2. Press the TRIGGER soft key.
3. Press the Trigger Cond. Setup soft key.
4. Press the Select soft key to select Enhanced.
Setting the Trigger Type
5. Press the Type soft key. The trigger type selection menu appears.
6. Use jog shuttle+SELECT to select Wave Window.
Setting the Trigger Conditions
7. Press the Set Pattern soft key. The Wave Window trigger setup menu appears.
8. Use jog shuttle+SELECT to select the pattern of each channel (CH1 to CH16)
from ON and OFF.
Note
• The wave window trigger does not apply to the 701265 (TEMP/HPV) and the 701280
(FREQ).
• Wave window trigger is not applicable when measuring temperature on the 701261
(UNIVERSAL) or 701262 (UNIVERSAL (AAF)).
• Wave window trigger cannot be used in Recorder mode.
Setting the Cycle Frequency, Reference Cycle, Synchronization Channel,
Level, and Hysteresis
8. Use jog shuttle+SELECT to set Cycle Frequency. Pressing RESET resets the
value to 50 Hz.
10. Likewise, set the Reference Cycle to 1, 2, or 4.
11. Likewise, set Sync. Ch to Auto or CH1 to CH16.
12. If you select a setting other than Auto in step 11, set the Level of the
synchronization channel.
13. Likewise, set Hys to
, , or .