7-8
IM 701240-01E
7.4 Acquiring Data Using the Sequential Store
Function (Single (N) Mode)
<For a description of this function, refer to page 2-29.>
Procedure
1. Press MENU.
2. Press the TRIGGER soft key.
3. Press the Mode soft key.
4. Press the Single(N) soft key to set the trigger mode to Single (N).
5. Press the Single(N) Count soft key.
6. Turn the jog shuttle to set Single(N) Count.
Explanation
You can use the sequential store function by setting the trigger mode to Single(N).
Acquisition Count (Single(N) Count)
Set a value between 1 and 2000.
However, the selectable acquisition count varies depending on the selected record
length and acquisition mode. For details, see appendix 3.
Displaying Waveforms
You can recall waveforms from memory in the same fashion as with the history memory
function. For details, see section 11.1.
Notes on When Using Sequential Store
If you stop the waveform acquisition by pressing START/STOP, sequential store also
stops. Sequential store restarts from the beginning when acquisition resumes.