App-32
IM 701240-01E
Appendix 14 High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS) (701260) Specifications
Item Specifications
Response time Rising (0 to 90% of 10 div): 100 ms (typical
10
)
(only when observing Falling (100 to 10% of 10 div): 250 ms (typical
10
)
RMS)
Bandwidth limit Select from OFF, 10 kHz, 1 kHz, and 100 Hz
Cutoff characteristics: –12 dB/OCT (typical
10
)
Probe attenuation setting Voltage probe: 1:1, 10:1, 100:1, and 1000:1
Current probe: 10 A:1 V (for the 701933) and 100 A:1 V (for the 701930/701931)
Compatible probes/cables Connection cable (for high voltage 1:1): Recommended 1
701901 (isolated type BNC-safety alligator clip adapter × 2: For measuring 850 V
(DC+ACpeak) or less),
701954 (alligator clip (dolphin type) red/black 2-piece set) is required separately
Voltage probe (10:1 safety probe): Recommended 2
700929 (10:1 safety probe) .20 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
Current probe (power can be supplied from the SL1400)
701930 (150 A), 701931 (500 A), and 701933 (30 A)
1 Value measured under standard operating conditions (section 18.11).
Withstand voltage: 3700 Vrms for 1 minute
Allowable transient surge voltage: ±5200 Vpea
k
(between earth and input)
Combined with the 700929
2
4
7
3
700929
Combined with the 701901+701954
H
L
Direct input (cable that does not comply with the safety standards)
8
9
BNC
H
L
701901
701954
5
H
L
6
10 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
When applying high voltage using this module, use the 1:1 safety cable
(combination of 701901 and 701954) or the isolated probe (700929).
Note that the overvoltage tolerance of the low and high sides of this module may
differ depending on the combination of the connected probe, measurement lead,
and clip.
Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
To prevent electric shock, be sure to furnish protective earth grounding of the
SL1400.
To prevent electric shock, be sure to fasten the module screws. Otherwise, the
electrical and mechanical protection functions will not be activated.
Avoid continuous connection under an environment in which the surge voltage
may occur.