7-13
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Waveform Acquisition
7.6 Setting the Action When Waveform Display Is
Updated (Action-on-Stop)
<For a description of this function, refer to page 2-32.>
Procedure
1. Press MENU.
2. Press the TRIGGER soft key.
3. Press the ACTION soft key.
Setting the Action
4. Press the Mode soft key. The Mode setup menu appears.
5. Press the On Stop soft key.
6. Press the Action soft key. The Action setup menu appears.
7. Use jog shuttle+SELECT to select the action to be enabled. (You can select
multiple actions.)
8. If you select Save to File in step 7, use the jog shuttle+SELECT to set the file
save format to Binary, ASCII, or Float. If you select Send Mail, use jog
shuttle+SELECT to set Mail Count.
Setting the Number of Actions
9. Press the Sequence soft key to select Single or Continue.
10. Press the ACQ Count soft key.
11. Turn the jog shuttle to set the number of actions. Pressing RESET resets the
count to Infinite.
Setting the Execution Mode of the Action-on-Stop at Power ON
12. Press MENU.
13. Press the Next 1/2 soft key.
14. Press the Others soft key.
15. Press the Action Mode (Power ON) soft key to select ON or OFF.
Executing Action-on-Stop
16. Press START/STOP. Waveform acquisition is started, and action-on-stop is
executed. While action-on-stop is in execution, is displayed in the upper left
of the screen. In addition, if the following action is specified, the following icon is
displayed at the upper left of the screen while the action is being executed.
Image/Save to File: Save destination medium (
PC card, SCSI device,
internal HD, network drive, or USB storage
device)
Send Mail:

Table of Contents