EasyManua.ls Logo

YOKOGAWA SL1400 - Connecting the Bridge Head

YOKOGAWA SL1400
589 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-27
IM 701240-01E
3
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Preparations for Measurements
3.11 Connecting the Bridge Head
Strain is measured by connecting a strain gauge bridge (bridge head) or a strain gauge
transducer to the strain module (701270 (STRAIN_NDIS) or 701271 (STRAIN_DSUB)).
This section will mainly describe the procedures and precautions related to the
connection of the bridge head (Model 701955/701956/701957/701958). For the
connection of other strain gauge bridges or strain gauge transducers, see the respective
manuals.
CAUTION
Only connect a strain gauge bridge (bridge head) or a strain gauge transducer
to the strain module. Connecting other devices or applying a voltage that
exceeds the values indicated below to the strain module may damage the input
section.
Maximum input voltage (between Input+ and Input–)
10 V (DC+ACpeak)
Maximum allowable common mode voltage (between each terminal and earth
ground)
42 V (DC+ACpeak) (CAT II, 30 Vrms)
Connecting the Strain Gauge
The bridge head (701955/701956/701957/701958) supports six types of connection
methods: single-gauge method, single-gauge three-wire method, adjacent-side two-
gauge method, opposite-side two-gauge method, opposite-side two-gauge three-wire
method, and four-gauge method. For details, see the manual that comes with the bridge
head (701955/701956/701957/701958).
If you are using a strain gauge bridge or a strain gauge transducer other than the bridge
head (701955/701956/701957/701958), see the respective manuals.
Connecting the Strain Module and the Bridge Head
When Using the Strain Module (701270) and the Bridge Head (701955/701956)
Using the cable that comes with the bridge head (701955/701956), connect the bridge
head to the 701270 (STRAIN_NDIS).
Strain Module
701270(STRAIN_NDIS)
Cable
Bridge head
701955/701956
When Using the Strain Module (701271) and the Bridge Head (701957/701958)
Using the cable that comes with the bridge head (701957/701958), connect the bridge
head to the 701271 (STRAIN_DSUB).
Cable
Bridge head
701957/701958
Strain Module
701271(STRAIN_DSUB)

Table of Contents

Related product manuals