Loading...
7-7
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Waveform Acquisition
Number of Data Points on Which Moving Average Is Computed
Sample Rate (S/s) Number of Data Points
10 M* 1 point
5 M 2 points
2 M 4 points out of 5 points
1 M 8 points out of 10 points
500 k 16 points out of 20 points
200 k 32 points out of 50 points
100 k 64 points out of 100 points
50 k 128 points out of 200 points
20 k 256 points out of 500 points
10 k or less 256 points out of (10 M ÷ sample rate)
* Same operation as the normal mode.
7.3 Acquiring Data Using Box Average