Loading...
6-7
IM 701240-01E
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Triggering
Selectable Range of the Hold Off Time
0 to 10 s (resolution: 0.1 µs)
Notes for Setting the Hold Off Time
If you are setting the hold off time to 50 ms or greater, set the trigger mode to normal.
When used with A->B(N) or A Delay B trigger, the hold off time applies only to
condition B.
The hold off time is set to 0 ns for Period trigger, and this function is invalid.
6.4 Setting the Hold Off Time