This function activates a trigger when the trigger source signal crosses a specified level.
Trigger Source
Select the trigger source from CH1 to CH16.
Note
If you set the trigger source to a value other than CH1 to CH16, see the sections below.
Time: Section 6.8, “Setting the Timer Trigger (SIMPLE)”
LogicA or Logic B: Section 6.9, “Setting the Logic Trigger (SIMPLE)”
Ext (TRIG IN terminal of the side panel): Section 6.6, “Setting the External Trigger (SIMPLE)”
Line (commercial power supply signal): Section 6.7, “Generating Triggers on the Power
Signal (SIMPLE)”
Trigger Level
When Measuring Voltage
Selectable range: Within voltage range
Resolution: Varies depending on the input module (see the table below).
When Measuring Temperature
Selectable range: Measurement range of each thermocouple
Resolution: 0.1°C or 0.1 K (set within the measurement range (varies
depending on the thermocouple type)
For a description of the measurement range of thermocouples,
see section 5.14.
When Measuring Strain
Selectable range: ±(Measurable range)
Resolution: 1 µSTR or 0.0005 mV/V
Trigger Level Resolution
The selectable resolution varies depending on the input module.
When Measuring Voltage (When Probe Attenuation Is 1:1)
Input Module
701250/701255/ 701251 701265 701267 701275
Voltage Range 701261/701262
1 mV 0.001 mV
2mV 0.002 mV
5 mV 0.005 mV
10 mV 0.01 mV 0.01 mV
20 mV 0.02 mV 0.02 mV
50 mV 0.05 mV 0.05 mV 0.05 mV 0.05 mV
100 mV 0.1 mV 0.1 mV 0.1 mV 0.1 mV
200 mV 0.2 mV 0.2 mV 0.2 mV 0.2 mV 0.2 mV
500 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV
1 V 0.001 V 0.001 V 0.001 V 0.001 V 0.001 V
2 V 0.002 V 0.002 V 0.002 V 0.002 V 0.002 V
5 V 0.005 V 0.005 V 0.005 V 0.005 V 0.005 V
10 V 0.01 V 0.01 V 0.01 V 0.01 V 0.01 V
20 V 0.02 V 0.02 V 0.02 V 0.02 V 0.02 V
50 V 0.05 V 0.05 V 0.05 V 0.05 V 0.05 V
100 V 0.1 V 0.1 V 0.1 V 0.1 V 0.1 V
200 V 0.2 V 0.2 V 0.2 V
500 V 0.5 V
1 kV 0.001 kV
2 kV 0.002 kV
The resolution is 0.01 division on all modules.