Triggering
Explanation
This function activates a trigger when the trigger source signal crosses a specified level.
Trigger Source
Select the trigger source from CH1 to CH16.
Note
If you set the trigger source to a value other than CH1 to CH16, see the sections below.
• Time: Section 6.8, “Setting the Timer Trigger (SIMPLE)”
• LogicA or Logic B: Section 6.9, “Setting the Logic Trigger (SIMPLE)”
• Ext (TRIG IN terminal of the side panel): Section 6.6, “Setting the External Trigger (SIMPLE)”
• Line (commercial power supply signal): Section 6.7, “Generating Triggers on the Power
Signal (SIMPLE)”
Trigger Level
When Measuring Voltage
Selectable range: Within voltage range
Resolution: Varies depending on the input module (see the table below).
When Measuring Temperature
Selectable range: Measurement range of each thermocouple
Resolution: 0.1°C or 0.1 K (set within the measurement range (varies
depending on the thermocouple type)
For a description of the measurement range of thermocouples,
see section 5.14.
When Measuring Strain
Selectable range: ±(Measurable range)
Resolution: 1 µSTR or 0.0005 mV/V
Trigger Level Resolution
The selectable resolution varies depending on the input module.
When Measuring Voltage (When Probe Attenuation Is 1:1)
Input Module
701250/701255/ 701251 701265 701267 701275
Voltage Range 701261/701262
1 mV – – 0.001 mV – –
2mV – – 0.002 mV – –
5 mV – – 0.005 mV – –
10 mV – 0.01 mV 0.01 mV – –
20 mV – 0.02 mV 0.02 mV – –
50 mV 0.05 mV 0.05 mV 0.05 mV – 0.05 mV
100 mV 0.1 mV 0.1 mV 0.1 mV – 0.1 mV
200 mV 0.2 mV 0.2 mV 0.2 mV 0.2 mV 0.2 mV
500 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV 0.5 mV
1 V 0.001 V 0.001 V 0.001 V 0.001 V 0.001 V
2 V 0.002 V 0.002 V 0.002 V 0.002 V 0.002 V
5 V 0.005 V 0.005 V 0.005 V 0.005 V 0.005 V
10 V 0.01 V 0.01 V 0.01 V 0.01 V 0.01 V
20 V 0.02 V 0.02 V 0.02 V 0.02 V 0.02 V
50 V 0.05 V 0.05 V 0.05 V 0.05 V 0.05 V
100 V 0.1 V 0.1 V 0.1 V 0.1 V 0.1 V
200 V 0.2 V 0.2 V – 0.2 V –
500 V – – – 0.5 V –
1 kV – – – 0.001 kV –
2 kV – – – 0.002 kV –
The resolution is 0.01 division on all modules.
6.5 Setting the Edge Trigger (SIMPLE)