4-19
IM 701240-01E
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
Common Operations
Settings after Auto Setup
Waveform acquisition and display conditions
START/STOP START
Acquisition mode Normal
Number of acquisitions Infinite
Time base Int
Accumulate mode OFF
Zoom target Channels whose display is turned ON
Vertical-axis settings
Voltage range Value that causes the absolute value of the input waveform
to be 1.6 to 4.5 divisions
Offset voltage 0 V or the center value between the maximum and
minimum values of the input signal
Coupling DC
Bandwidth limit FULL
Display ON/OFF Turns ON channels whose absolute value of the input
voltage is greater than or equal to 20 mV (1:1)
Position 0 division
Variable mode OFF
V Zoom ×1
Scaling OFF
Horizontal-axis settings
Record time Value that the displays 1.6 to 4 periods of the waveform
with the shortest period of the auto setup target waveforms
Trigger settings
Trigger mode Auto
Trigger type SIMPLE
Trigger source Channel whose waveform amplitude is greater than or
equal to 1 division with the longest period
Trigger level/slope Center level between the maximum and minimum values/
rising
Trigger coupling DC
Hysteresis (high sensitivity)
Hold off time 0.0 ns
Trigger position 50.0%
Trigger delay 0.0 µs
Computation settings
Math OFF
Other settings
Realtime output OFF
Accumulation OFF
Logic A and Logic B display OFF
Items other than those listed above are not applicable for auto setup.
Note
If none of the input waveforms are detected on the target module, the waveform acquisition is
not started.
4.5 Performing Auto Setup