6-2
IM 701240-01E
Single(N) Mode
When the trigger condition is met, the waveform is acquired the specified number of
times and stops. After waveform acquisition is finished, all the waveforms are displayed.
This mode is used when acquiring a waveform using the sequential store function.
Log Mode
In this mode, the trigger settings are disabled. The specified record length of data is
acquired once when acquisition is started, and the displayed waveforms are updated. If
the time axis setting is in the range in which the display mode is set to roll mode, the
display is set to roll mode.
Note
• The trigger mode setting applies to both simple and enhanced triggers.
• In Recorder mode (see chapter 9), select the trigger mode from Auto, Repeat, Single, and
Log. For details on the Repeat mode, see section 9.2.
6.1 Setting the Trigger Mode