EasyManua.ls Logo

YOKOGAWA SL1400 - Page 210

YOKOGAWA SL1400
589 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-22
IM 701240-01E
Explanation
This function activates a trigger the first time condition B becomes true after condition A
has become true and a preset time has elapsed.
Condition A and Condition B (Set Pattern)
Pattern of each channel: CH1 to CH16, Logic A, and Logic B
Select from the following:
CH1 to CH16 (Other Than Logic Inputs)
H: Above the preset trigger level
L: Below the preset trigger level
X: Don’t Care
Logic Input
Enable: Make the combination of the pattern
1
of each bit the trigger condition
Disable: Don’t Care
1 Select the pattern of each pattern from the following:
H: Above a certain level
2
L: Below a certain level
2
X: Don’t Care
2 Varies depending on the logic probe being used as follows:
702911/702912/700986: Approx. 1.4 V
700987: 6 V ± 50% (for DC input)
700987: 50 V ± 50% (for AC input)
Condition
Select from the following:
Enter: A trigger is activated when all channels match the specified pattern.
Exit: A trigger is activated when any of the channels no longer match the specified
pattern.
Delay Time
0 to 10 s (resolution is 100 ns)
Trigger Level
The trigger level setting applies to both simple and enhanced triggers.
See “Trigger Level” in section 6.5.
Trigger Hysteresis
Sets a width to the trigger level so that triggers are not activated by small changes in the
trigger signal. Select the trigger hysteresis from , , and .
See “Trigger Hysteresis” in section 6.5.
Hold Off
See section 6.4.
Note
If you want to activate only one pattern trigger, use the A->B(n) trigger function (see section
6.10).
6.11 Setting the A Delay B Trigger (ENHANCED)