6-32
IM 701240-01E
Explanation
The following three pulse width triggers are available.
B > Time: A trigger is activated when condition B goes false after holding true for the
preset pulse width.
B < Time: A trigger is activated when condition B goes false after holding true for less
than the preset pulse width.
B Time Out: A trigger is activated when condition B has held true for the preset pulse
width.
The point where the trigger is activated differs between B > Time and B Time Out as
shown in the figure below.
When B > Time, CH1 = H, and Time = 4 µs
Trigger
5
µ
s
1
µ
s
CH1
Trigger
5
µ
s
1
µ
s
CH1
4
µ
s
When B Time Out, CH1 = H, and Time = 4 µs
B<Time Setup Example
When B < Time, condition B: CH1 = H, CH2 = H, and Time = 4 µs
Trigger
LHL
HHLLL
H
CH1
CH2
CH1
CH2
3
µ
s 5
µ
s
3
µ
s
6.14 Setting the B > Time, B < Time, or B Timeout (Pulse Width) Trigger (ENHANCED)