11-12
IM 701240-01E
For V (Vertical) Cursors (When Not Displaying the X-Y Waveform)
1. Press CURSOR.
Selecting the Cursor Type
2. Press the Type soft key. The cursor type selection menu appears.
3. Press the Vertical soft key.
Selecting the Waveform to Be Measured
4. Press the Trace soft key. The trace selection menu appears.
5. Use jog shuttle+SELECT to select the waveform to be measured.
If you set the waveform to be measured to All, LogicA, LogicB, or LogicA &
LogicB, proceed to step 6. If not, proceed to step 9.
Setting the Logic (When Trace Is Set to All, LogicA, LogicB, or LogicA & LogicB)
6. Press the Logic Setup soft key. The logic setup menu appears.
Selecting the Notation of Cursor Measurement Values
7. Use jog shuttle+SELECT to set Format to Binary or Hexa.
Note
The notation selected in step 7 applies to the notation of the numeric monitor display that is
shown on the right side of the screen. For the procedure of displaying the numeric monitor,
see section 8.13.
Selecting the Read Direction of Bit Data
8. Use jog shuttle+SELECT to set Order to A1->A8 or B1->B8.
Moving the Cursor
9. Press the Cursor1/Cursor2 soft key to set the jog shuttle control to Cursor1.
10. Turn the jog shuttle to move Cursor1.
11. Likewise, move Cursor2. If the jog shuttle control is set to both Cursor1 and
Cursor2, both cursors are moved.
Logic Setup appears
only when Trace is set
to the following:
All, LogicA, LogicB, or
LogicA & LogicB
11.4 Measuring Waveforms Using Cursors