2-14
IM 701240-01E
Duty Cycle
Duty cycle (%) = T
high
(s)/T
w
(s) or Duty (%) = T
low
(s)/T
w
(s)
T
high
(s)
T
w
(s)
T
low
(s)
Pulse Width
Pulse width (s) = T
high
(s) or Pulse width (s) = T
low
(s)
T
high
(s)
T
low
(s)
Power Supply Frequency
Power supply frequency (Hz) = 1/T
w
(s)
T
w
(s)
Pulse Integration (Distance/Flow Rate)
Pulse integrated value = N (count) × physical amount per pulse (I)
Set the physical amount per pulse (I) to distance or flow rate. A suitable user-defined
unit can be assigned to the specified physical amount.
N (count)
l
Velocity
Velocity (km/h) = Distance per pulse l (km)/T
w
(s) × 3600
Velocity (m/s) = Distance per pulse l (m)/T
w
(s)
The distance and unit can be user defined (angular velocity, etc.).
T
w
(s)
Distance
per pulse (l)