2-18
IM 701240-01E
Enhanced Trigger
A >B(N) Trigger
A trigger is activated the N
th
time condition B becomes true after condition A has
become true.
A
B
True
True N times
Trigger
Condition A
Condition B
Condition
Enter: True when the preset condition is met
Exit: True when the preset condition is not met
LogicA/LogicB
Disable: Do not use logic input
Enable: Use logic input
CH1 to CH16
L: Low level
H: High level
X: Don't care
A Delay B Trigger
A trigger is activated the first time condition B becomes true after condition A has
become true and a preset time has elapsed.
Condition A
Condition B
Condition
Enter: True when the preset condition is met
Exit: True when the preset condition is not met
LogicA/LogicB
Disable: Do not use logic input
Enable: Use logic input
CH1 to CH16
L: Low level
H: High level
X: Don't care