17-11
IM 701240-01E
17
Troubleshooting, Maintenance, and Inspection
Code Message and Corrective Action Section
1313 FTP Error: AppendData Chapter 15
Confirm the network settings, connection, and disk capacity.
1314 FTP Error: Client Handle Chapter 15
Confirm the network settings and connection.
1315 FTP Error: Others Chapter 15
Confirm the network settings and connection.
1335 Cannot send data to a network printer. 12.3, 15.4
Confirm the network settings and connection.
1336 Cannot send a mail. 15.5
Confirm the network settings and connection.
1345 Test Error. Chapter 15
1346 Test Success. Chapter 15
1348 Failed to initialize network. Chapter 15
Confirm the network settings.
1349 The password entered the first time is different from the password entered the second time. 15.6
Reenter the password for the second time.
1350 Failed to acquire time from SNTP server. 15.8
Confirm the network settings and connection.
1501 To apply the changes, power-cycle the SL1400. Chapter 15
1502 Executed the firmware overwriting of the frequency module. 17.4
1503 Overwriting firmware of the frequency module... 17.4
1505 Executed the firmware overwriting of the built-in printer. 17.4
1506 Overwriting the printer firmware... –
1600 This function can be used only when the recorder mode is Off. Chapter 9
1601 Can be specified only when the recorder mode is Off. Chapter 9
1602 Can be configured only when the recorder mode is Off. Chapter 9
Print from the Reprint menu.
1603 This function cannot be used when the recorder mode is set to X-Y. 9.6
1604 Cannot be configured or executed when the recorder mode is Off. Chapter 9
1605 Repeat trigger cannot be specified. Chapter 9
Repeat trigger is valid only during chart recorder mode when the print style is waveform and shot
recording is not OFF.
1606 Cannot be changed during X-Y recording. –
1607 Cannot use wave window trigger when the recorder mode is set to chart or X-Y. 6.17
1608 Can not be specified when the print style is Numeric. 9.5
1611 Cannot be specified because characters in the JIS level-2 kanji set are included. 4.2
Create the file on the local drive, and then copy it to the network drive.
1612 Can not be specified in the current mode. –
Other Errors (9999)
Code Message and Corrective Action Section
9999 This error No. is not defined. –
Note
If servicing is required, initialize the instrument once for confirmation.
17.2 Messages and Corrective Actions