2-20
IM 701240-01E
B > Time Trigger, B < Time Trigger, and B TimeOut Trigger
A trigger is activated on the falling or rising edge of the pulse when the pulse width
(width over which condition B is met) exceeds (or drops below) a preset time. In the
case of a timeout trigger, a trigger is activated when a preset time elapses.
LogicA/LogicB
Disable: Do not use logic input
Enable: Use logic input
CH1 to CH16
L: Low level
H: High level
X: Don’t care
•B > Time Trigger
B
Trigger
Time
B: Pulse width
•B < Time Trigger
B
Trigger
Time
•B TimeOut Trigger
B
Trigger
Time
2.3 Setting the Trigger