Item Specifications
Safety standard Complying standard
1
EN61010-1, EN61010-2-030, EN61010-031
Overvoltage Category (Installation Category) II
2
Measurement Category: See the specifications of each module
3
Pollution degree 2
4
Already certified (701240/701250/701251/701255/701267/701261/701262/
701265/701270/701271/701275/701280/700986/700987/701955/701956/
701957/701958)
Emission Complying standard EN61326-1 Class A
EN61326-2-1
EN 55011 Class A, Group 1
EMC Regulatory Arrangement in Australia and New Zealand EN 55011
Class A, Group 1, Korea Electromagnetic Conformity Standard
( )
Already certified (701240/701250/701251/701255/701267/701261/701262/
701265/701270/701271/701275/701280/700986/700987/701955/701956/
701957/701958)
EN61000-3-2
EN61000-3-3
This product is a Class A (for industrial environment) product.
Operation of this product in a residential area may cause radio
interference in which case the user is required to correct the interference.
Cable condition Shared external trigger/external clock input terminal
Use the BNC-RCA adapter (YOKOGAWA: 366928) and a BNC cable
5
and attach a ferrite core (TDK: ZCAT2035-0930A, YOKOGAWA:
A1190MN) on one end (SL1400 end).
Trigger output terminal
Same as the external trigger input terminal above.
Video signal output (VIDEO OUT (SVGA)) terminal
Use a 15-pin D-Sub VGA shielded cable
5
.
GP-IB interface connector
Use shielded GP-IB cables
5
.
Serial (RS-232) interface connector
Use a shielded RS-232 cable
5
and attach a ferrite core (TDK: ZCAT2035-
0930A, YOKOGAWA: A1190MN) on one end (SL1400 end).
SCSI connector
Use a shielded SCSI cable
5
and attach a ferrite core (TDK: ZCAT2035-
0930A, YOKOGAWA: A1190MN) on one end (SL1400 end).
USB peripheral connector
Use shielded USB cables.
5
USB interface connector
When connecting a USB keyboard or mouse using a shielded cable or
when connecting to a USB printer, use shielded USB cables.
5
Ethernet connector
Use category 5 Ethernet cables
6
or better cables.
Probes connected to modules and wiring
Use twisted pair cables when connecting items other than probes to the
module (excluding the 701267). Attach a ferrite core (TDK: ZCAT2035-
0930A, YOKOGAWA: A1190MN) to the probes and cables that are
connected to the modules. Wrap the cable around the ferrite core once.
For the 701267
Bundle the H and L measurement leads 758933. Attach a ferrite core
(TDK: ZCAT2035-0930AÅCYOKOGAWA: A1190MN) to the SL1400
side of the bundled leads.
Logic probe input
Attach a ferrite core (TDK: ZCAT2035-0930A, YOKOGAWA: A1190MN)
on one end (SL1400 end) of the cable to be connected to the logic probe
input terminal.
Current probe (701930, 701931, and 701933)
When connecting a current probe to the input terminal and probe power
terminal of the module, attach a ferrite core (TDK: ZCAT2035-0930A,
YOKOGAWA: A1190MN) on one end (SL1400 end) of the two cables
together.
Bridge head for the strain module
When using the 701270, use the 701955 or the 701956; when using the
701271, use the 701957 or the 701958.
18.11 General Specifications