18-20
IM 701240-01E
WARNING
Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
To prevent electric shock, be sure to furnish protective earth grounding of the
SL1400.
To prevent electric shock, be sure to fasten the module screws.
The module screws must be fastened for the module to function as a non-
isolation module. In addition, all electrical and mechanical protection functions
are activated only when the screws are fastened.
The maximum input voltage of the module is valid only when all the screws are
fastened, and the protection path of the metal BNC is secured.
18.12 Module Specifications