18-22
IM 701240-01E
Withstand voltage: 3700 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±5200 Vpeak
H
L
2
4
(Red)
(Black)
758933 701954
3
H
L
5
6
With the 758933 and 701954
Input
terminal
(Red)
(Black)
Input
terminal
Direct input (cable that does not
comply with the safety standards)
7 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
When measuring high voltage using this module, use the 1:1 safety cable
(combination of 701901 and 701954) or the isolated probe (700929).
The measurement category of this module when used with 758933 measurement
leads and 701954 alligator clips is 400V-CAT II on the low side and 700V-CAT II
on the high side. Be careful because the overvoltage tolerance between the low
and high sides is different.
Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
To prevent electric shock, be sure to furnish protective earth grounding of the
SL1400.
To prevent electric shock, be sure to fasten the module screws. Otherwise, the
electrical and mechanical protection functions will not be activated.
Avoid continuous connection under an environment in which the surge voltage
voltage may occur.
18.12 Module Specifications