Item Specifications
Table of cutoff frequency characteristics of the anti-aliasing filter (AAF)
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample rate AAF Low-Pass Filter
100 kS/s 40 kHz OFF
50 kS/s 20 kHz OFF
20 kS/s 8 kHz OFF
10 kS/s 4 kHz 4 kHz
5 kS/s 2 kHz 4 kHz
2 kS/s 800 Hz 4 kHz
1 kS/s 400 Hz 400 Hz
500 S/s 200 Hz 400 Hz
200 S/s 80 Hz 400 Hz
100 S/s 40 Hz 40 Hz
50 S/s 20 Hz 40 Hz
20 to 5 S/s 20 Hz 40 Hz
2 S/s or less 20 Hz 40 Hz
Ext sample 40 kHz OFF
1 Value measured under standard operating conditions (section 18.11).
H
L
2
3
BNC
Combined with the 10:1 passive probe
(701940)
Direct input (cable that does not comply
with the safety standards)
H
L
2
3
4 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
• Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
• To prevent electric shock, be sure to furnish protective earth grounding of the
SL1400.
• To prevent electric shock, be sure to fasten the module screws. Otherwise, the
electrical and mechanical protection functions will not be activated.
• Avoid continuous connection under an environment in which the surge voltage
may occur.
18.12 Module Specifications