18-34
IM 701240-01E
Item Specifications
Input voltage range (±FS)1:1 ProbeFactor: ±1 V, ±2 V, ±5 V, ±10 V, ±20 V, and ±50 V(±FS)
10:1 ProbeFactor: ±10 V, ±20 V, ±50 V, ±100 V, ±200 V, 500 V(±FS)
(combined with the recommended probe 700929)
Input impedance 1 M±1%, approx. 35 pF
Pull-up function: 4.7 k, approx. 5 V (pull-up can be turned ON only when the input is set to Pull-Up
5V)
Input coupling AC and DC
Probe attenuation setting 10:1 and 1:1
Minimum voltage width 200 mV
P-P
for pulse detection
Bandwidth limit Select from Full, 100 kHz, 10 kHz, 1 kHz, and 100 Hz
Cutoff characteristics: –12 dB/OCT (typical
9
)
Threshold Set within the FS of the voltage range. Set in units of 1% of the FS.
Hysteresis Select ±1%, ±2.5%, or ±5% of the FS of the voltage range
Preset function Logic (5V/3V/12V/24V), electromagnetic pickup, zero crossing, pull-up, AC100V, AC200V, and
user-defined
Slope selection Select rising or falling
Lower –3 dB point when 0.5 Hz or less (0.05 Hz or less when using the 700929) (typical
9
)
AC coupled
Chatter elimination function OFF or 1 to 1000 ms (1 ms resolution)
Eliminates the chatter that occurs such when the contact input is turned ON/OFF. Can discard the
signal changes over the specified interval.
Input status indication Input status indication through the LEDs of each channel
function When in operation: Illuminates in green when pulse input is detected
When overdriven: Illuminates in red when the input voltage exceeds the range
Compatible probes/cables Connection cable (1:1): Recommended 1
366926
Voltage probe (10:1 safety probe): Recommended 2
700929 (10:1 safety probe) 20 to 45 pF: For measuring 1000 V (DC+ACpeak) or less
1 Value measured under standard operating conditions (section 18.11).
H
L
Combined with the 700929
2
700929
H
L
Combined with the (701901 + 701954) or
direct input (cable that does not comply with the safety standards)
4
5
BNC
Withstand voltage: 1500 Vrms for 1 minute
Allowable transient surge voltage (between earth and input): ±2100 Vpeak
3
6 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 1 µs (input range: ±10 V, bandwidth limit: Full, and
hysteresis: ±1%)
7 Input waveform of 90 Vrms, sine wave (input range: AC100V, bandwidth limit 100 kHz, and hysteresis: ±1%)
8 Input waveform of 1 Vpp, rectangular wave, rise/fall time within 5 ns (input range: ±10 V, bandwidth limit: Full, and hysteresis: ±1%)
9 Typical value represents a typical or average value. It is not strictly warranted.
WARNING
Do not apply input voltage exceeding the maximum input voltage, withstand
voltage, or allowable surge voltage.
To prevent electric shock, be sure to furnish protective earth grounding of the SL1400.
To prevent electric shock, be sure to fasten the module screws. Otherwise, the
electrical and mechanical protection functions will not be activated.
Avoid continuous connection under an environment in which the surge voltage
may occur.