AC100V
Setup Item Setting
V Range ±200 V
Coupling AC
Probe 10:1
Bandwidth Yes
2
(Full is not allowed)
Threshold 0V
Hys Yes
3
Slope Rising edge
Chatter Elimination Yes
5
Pull Up No
AC200V
Setup Item Setting
V Range ±500 V
Coupling AC
Probe 10:1
Bandwidth Yes
2
(Full is not allowed)
Threshold 0 V
Hys Yes
3
Slope Rising edge
Chatter Elimination Yes
5
Pull Up No
EM Pickup
Setup Item Setting
V Range ±1 V
Coupling DC
Probe 1:1
Bandwidth Yes
2
Threshold 0 V
Hys Yes
3
Slope Rising edge
Chatter Elimination Yes
5
Pull Up No
User
Setup Item Setting
V Range Yes
7
Coupling Yes
8
Probe Yes
1
Bandwidth Yes
2
Threshold Yes
9
Hys Yes
3
Slope Yes
4
Chatter Elimination Yes
5
Pull Up No
When you select a preset, the setup items are automatically set to the settings in the table. The meaning of Yes and No in the table
is as follows:
Yes: Item that can be set to an arbitrary value
No: Item that cannot be set (not displayed on the menu)
1 Probe type: Select 1:1 or 10:1.
2 Bandwidth limit: Select 100 Hz, 1 kHz, 10 kHz, 100 kHz, or Full. Full is not selectable for AC100V and AC200V.
3 Hysteresis: Select ±1%, ±2.5%, or ±5%.
4 Slope: Select rising or falling.
5 Chatter elimination: Select from 0 ms to 1000 ms.
6 Pull-up: Select ON or OFF. Selectable only for Pull-up 5V.
7 Voltage range:
(Probe = 1:1) Select ±1 V, ±2 V, ±5 V, ±10 V, ±20 V, or ±50 V.
(Probe = 10:1) Select ±10 V, ±20 V, ±50 V, ±100 V, ±200 V, or ±500 V.
8 Coupling: Select DC or AC.
9 Threshold level: Set within the specified voltage range.
Appendix 12 List of Preset Settings of the Frequency Module