AutoTest Sytem Operation
Subject to Export Control, see Cover Page for details.
5-10
(System) Command Elements
Commands in this group are system specific but adhere to the following guidelines. Refer to the TETRA
Option Manual for details on each TETRA System's AUTOTEST commands.
Setting Values
set_parameter_name value1 [ value_x … ]
Sets specified parameter to the value(s). Values may be numeric or strings.
set_parameter_name
{ value1 [ value_x … ] } { value1 [ value_x … ] } { value1 [ value_x … ] } . . .
Sets specified parameter to the value(s) within the { } specified by the n value of repeat n.
Values may be numeric or strings. If nth value(s) not in list, use last value(s).
Tests: of protocol, measurements, . . .
test_test_name [ parameter_value(s) … ]
Performs the named test using the supplied parameter(s) (if necessary). Returns a pass | fail
flag.
Common Commands
The following is a list of Common Commands used throughout the 3900 AutoTest System.
Do not include brackets ( [ ] ) in command script.
send_date_time
Sends “HPW_TOD DD/MM/YYYY hh:mm:ss” of the 3900 at the time the command is sent to the
radio.
where:
DD = Day
MM = Month
YYYY = Year
hh = Hour (in 24 hour format)
mm = Minutes
ss = seconds
sw_pause
Pauses script. Requires user prompt to Abort or Continue test script.
set_analog_timeout
Sets Test Set’s analog timeout setting.
Format: Integer value in seconds.
set_ptt_out
Enables/Disables PTT function.
Valid states are:
0 = Disabled
1 = Enabled
set_rfgen_freq [ frequency in MHz ]
Sets generator frequency and sends the rfgen setfreq command to generator sub-system.
Format: Decimal frequency value in MHz