3900 Platform Specifications
Subject to Export Control, see Cover Page for details.
B-25
Power Requirements
AC
Voltage
100 V to 120 VAC @ 60 Hz
220 V to 240 VAC @ 50 Hz
Power Consumption
Nominally 120 W (200 W Max)
Mains Supply Voltage Fluctuations
< 10% of the nominal voltage
Fuse Requirements
3 A, 250 V, Type F
Environmental
Operating Temperature
0 to 50° C (Tested in accordance with MIL-PRF-28800F Class 3)
Warm-up Time
15 minutes
Storage Temperature
-40 to 71°C (Tested in accordance with MIL-PRF-28800F Class 3)
Relative Humidity
80% up to 31°C decreasingly linearly to 50% at 40°C
(Tested in accordance with MIL-PRF-28800F Class 3)
Altitude
4,000 m (13,123 ft) (Tested in accordance with MIL-PRF-28800F Class 3)
Shock and Vibrations
30 G Shock (Functional Shock)
5-500 Hz random vibrations
(Tested in accordance with MIL-PRF-28800F Class 3)
Use
Pollution Degree 2
EMC
3920
EN 61326, Class A
Reliability
> 8,000 hour calculated MTBF (MIL-HDBK-217F, Notice 2)
Safety Standards
3901/3902
Ul 61010B-1, EN 61010-1, CSA C22.2 No. 61010-1
3920
UL 61010-1