3900 Platform Specifications
Subject to Export Control, see Cover Page for details.
B-6
RF Receive Measurements (cont)
Demodulation
Demod Counter
Frequency
Range
20 Hz to 20 kHz (1 to 100 kHz FM Deviation, IF BW set appropriately for the received
modulation BW)
20 Hz to 10 kHz (30% to 90% AM, IF BW set appropriately for the received modulation BW).
Resolution
0.1 Hz
Accuracy
±50 ppm (±10 ppm Typical)
Input Waveform
Sine or Square
RF Characteristics
Input RF (Usable from 100 kHz)
10 MHz to 1.05 GHz (3901, 3920 Standard)
10 MHz to 2.7 GHz (3902, 3920 Freq Extension Option (390XOPT058))
RF Level
T/R: -10 to 50 dBm
ANT: -80 to +10 dBm
FM Deviation Meter
Range
0 to 150 kHz
Scales
1 to 200 kHz in a 1, 2, 5 sequence, plus Autoscale
Resolution
10 Hz
Accuracy
±3% plus source residual, ±1 count (1 to 150 kHz FM deviation, IF BW set appropriately for the
received modulation BW)
FM
Rate
20 Hz to 20 kHz (IF BW set appropriately for the received modulation BW)
RF Range (Usable from 100 kHz)
10 MHz to 1.05 GHz (3901, 3920 Standard)
10 MHz to 2.7 GHz (3902, 3920 Freq Extension Option (390XOPT058))
RF Level
T/R: -10 to +50 dBm
ANT: -80 to +10 dBm