EasyManua.ls Logo

Aeroflex 3901 - Page 290

Aeroflex 3901
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix G
Subject to Export Control, see Cover Page for details.
G-2
Error Number -113
"Undefined header"
Issued Remote Command does not define action to be performed.
i.e., :SA:TRAC:AVER:VAL command must include “?” to define command as a query command
or include a numeric value to define command as a set command.
Error Number -121
"Invalid character in number"
Character entered is not valid for selected parameter.
Error Number -123
"Exponent too large"
Number of digits entered exceeds maximum allowed.
Error Number -124
"Too many digits"
Number of digits entered exceeds maximum allowed.
Error Number -128
"Numeric data not allowed"
Selected parameter requires text data.
Error Number -131
"Invalid suffix"
Invalid unit of measurement has been defined in relation to specified measurement.
i.e., specifying dBm as the unit of measurement for a frequency parameter.
Error Number -141
"Invalid character data"
Text entered is invalid for selected parameter.
Error Number -144
"Character data too long"
Text stream exceeds maximum allowed for selected parameter.
Error Number -148
"Character data not allowed"
Selected parameter only accepts numeric entries.
Error Number -151
"Invalid string data"
Remote Command contains error in data string.
i.e., :TBS:CONF:CHPL:NEW plan_name contains more than 20 characters permitted.
Error Number -158
"String data not allowed"
Remote Command contains parameters not permitted in command script.
i.e., :TBS:CONF:CHPL:NEW defining Block 2 data when Block 2 has not been included (INCL) in
Channel Plan.