Chapter 10 157
Command Reference
:CORR:COLL
10. Command Reference
:CORR:COLL
Syntax [:SENSe]:CORRection:COLLect[:ACQuire] {STANdard1|STANdard2|STANdard3}
Description Measures the compensation data for OPEN/SHORT/LOAD compensation and turns on the
compensation function.
These compensation data are measured for the measurement frequency used in executing
the command (set with the :SOUR:FREQ command).
When using the multi-compensation function (set to ON with the :CORR:MULT
command), this is measured as the compensation data for the selected channel when
executing the command (selected with the :CORR:MULT:CHAN command). (No query)
Parameters
Related commands :CORR:OPEN on page 165
:CORR:SHOR on page 165
:CORR:LOAD on page 160
:CORR:MULT on page 161
:CORR:MULT:CHAN on page 161
Equivalent key
sequence
[Open]( [Blue] - [4] ) - OpenMeas
[Short]( [Blue] - [5] ) - ShortMeas
[Load]( [Blue] - [6] ) - LoadMeas
Description
STANdard1 Specifies the OPEN compensation.
STANdard2 Specifies the SHORT compensation.
STANdard3 Specifies the LOAD compensation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals