Chapter 10 189
Command Reference
:STAT:OPER?
10. Command Reference
:STAT:OPER?
Syntax :STATus:OPERation[:EVENt]?
Description Reads out the value of the Operation Status Event register. (Query only)
Query response {numeric}<newline><^END>
Related commands *CLS on page 124
Equivalent key
sequence
No equivalent key is available on the front panel.
:STAT:OPER:COND?
Syntax :STATus:OPERation:CONDition?
Description Reads out the value of the Operation Status Condition register. (Query only)
Query response {numeric}<newline><^END>
Equivalent key
sequence
No equivalent key is available on the front panel.
:STAT:OPER:ENAB
Syntax :STATus:OPERation:ENABle <numeric>
:STATus:OPERation:ENABle?
Description Sets the value of the Operation Status Enable register.
Parameters
If the specified parameter is out of the allowable setup range, the result of bit-by-bit logical
AND with 32767 (0x7fff) is set.
Query response {numeric}<newline><^END>
Related commands *SRE on page 127
:STAT:PRES on page 190
Equivalent key
sequence
No equivalent key is available on the front panel.
<numeric>
Description The value of the Enable register.
Range 0 to 32767
Initial value 0
Resolution 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals