Chapter 10 183
Command Reference
:INIT
10. Command Reference
:INIT
Syntax :INITiate[:IMMediate]
Description Starts the trigger system once.
If the trigger system is in the idle state, executing this command starts the trigger system
and, after measurement is executed once, returns the instrument to the idle state.
If the trigger system is not in the idle state or successive startup of the trigger system is
enabled (set to ON with the :INIT:CONT command), executing this command causes an
error and the command is ignored.
For details on the trigger system, refer to “Trigger system” on page 52. (No query)
Related commands :INIT:CONT on page 183
Equivalent key
sequence
No equivalent key is available on the front panel.
:INIT:CONT
Syntax :INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description Determines whether to start up the trigger system successively.
For details on the trigger system, refer to “Trigger system” on page 52.
Parameters
This setup is initialized to ON when using the :SYST:PRES command and to OFF when
using the *RST command.
Query response {1|0}<newline><^END>
Related commands *RST on page 126
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1 (initial value) Enables successive startup.
OFF or 0 Disables successive startup.