EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>4288A

Agilent Technologies 4288A User Manual

Agilent Technologies 4288A
289 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
Chapter 10 133
Command Reference
:CAL:CABL
10. Command Reference
:CAL:CABL
Syntax :CALibration:CABLe <numeric>
:CALibration:CABLe?
Description Sets the measurement cable length.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Cable]( [Blue] - [2] )
<numeric>
Description Desired cable length
Range 0 to 2
Initial value 0
Unit m (meter)
Resolution 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals