176 Chapter 10
Command Reference
:DISP
:DISP
Syntax :DISPlay[:WINDow][:STATe] {ON|OFF|1|0}
:DISPlay[:WINDow][:STATe]?
Description Enables/disables the display of the measurement result.
If you disables the display, “DISP OFF” is always displayed on the screen.
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[Disp Mode]( [Blue] - [Freq] ) - On/Off
:DISP:TEXT1:DIG
Syntax :DISPlay[:WINDow]:TEXT1[:DATA]:DIGit <numeric>
:DISPlay[:WINDow]:TEXT1[:DATA]:DIGit?
Description Specifies the number of displayed digits for both the primary and secondary parameters.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Disp Mode]( [Blue] - [Freq] ) - Digit
Description
ON or 1 (initial value) Enables the display.
OFF or 0 Disables the display.
<numeric>
Description Number of displayed digits for the primary/secondary parameter
Range 4 to 6
Initial value 6
Unit Digit
Resolution 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals