Chapter 10 185
Command Reference
:RANG:AUTO
10. Command Reference
:RANG:AUTO
Syntax [:SENSe][:FIMPedance]:RANGe:AUTO {ON|OFF|1|0}
[:SENSe][:FIMPedance]:RANGe:AUTO?
Description Selects the measurement range mode from the auto range (automatic range switching) or
hold range (fixed range).
Setting the measurement range (set with the :RANG command) automatically selects the
hold range mode.
Parameters
Query response {1|0}<newline><^END>
Related commands :RANG on page 184
Equivalent key
sequence
[Auto/Hold]
Description
ON or 1 (initial value) Specifies the auto range mode.
OFF or 0 Specifies the hold range mode.