174 Chapter 10
Command Reference
:DATA:FEED:CONT
:DATA:FEED:CONT
Syntax :DATA:FEED:CONTrol {BUF1|BUF2|BUF3},{ALWays|NEVer}
:DATA:FEED:CONTrol? {BUF1|BUF2|BUF3}
Description Determines whether to feed the measurement data into data buffer 1, data buffer 2, or data
buffer 3.
Parameters
Query response {ALW|NEV}<newline><^END>
Related commands :DATA:FEED on page 173
:DATA:POIN on page 175
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
BUF1 Specifies data buffer 1 as the buffer to which the
{ALWays|NEVer} setup is applied or reads out the setup of data
buffer 1.
BUF2 Specifies data buffer 2 as the buffer to which the
{ALWays|NEVer} setup is applied or reads out the setup of data
buffer 2.
BUF3 Specifies data buffer 3 as the buffer to which the
{ALWays|NEVer} setup is applied or reads out the setup of data
buffer 3.
Description
ALWays Feeds the measurement data into the data buffer specified with
{BUF1|BUF2|BUF3} each time a measurement is performed.
NEVer (initial
value)
Does not feed the measurement data into the data buffer specified
with {BUF1|BUF2|BUF3}.