Chapter 10 175
Command Reference
:DATA:POIN
10. Command Reference
:DATA:POIN
Syntax :DATA:POINts {BUF1|BUF2|BUF3},<numeric>
:DATA:POINts? {BUF1|BUF2|BUF3}
Description Specifies the size of data buffer 1, data buffer 2, or data buffer 3 in number of
measurements. Executing this command returns the pointer to the specified data buffer (the
location to feed measurement data) back to the start.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Related commands :DATA:FEED on page 173
:DATA:FEED:CONT on page 174
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
BUF1 Specifies data buffer 1 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 1.
BUF2 Specifies data buffer 2 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 2.
BUF3 Specifies data buffer 3 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 3.
<numeric>
Description Number of measurements for the data buffer specified with
{BUF1|BUF2|BUF3}
Range Data buffer 1: 1 to 200
Data buffer 2: 1 to 200
Data buffer 3: 1 to 1000
Initial value Data buffer 1: 200
Data buffer 2: 200
Data buffer 3: 1000
Resolution 1