Chapter 10 141
Command Reference
:CALC:COMP:MODE
10. Command Reference
:CALC:COMP:MODE
Syntax :CALCulate:COMParator:MODE {ABS|DEV|PCNT}
:CALCulate:COMParator:MODE?
Description Determines how to specify the limit range of the primary parameter for the comparator
function.
Parameters
Query response {ABS|DEV|PCNT}<newline><^END>
Related commands :CALC:COMP:PRIM:NOM on page 144
:CALC:COMP:PRIM:BIN{1-9} on page 142
Equivalent key
sequence
[Comprtr]( [Blue] - [1] ) - Limit - ∆ Mode
Description
ABS
(initial value)
Specifies the limit border value in an absolute value (absolute mode).
DEV
Specifies the border value relative to the reference value
*1
as an
absolute value (absolute tolerance mode).
PCNT Specifies the border value relative to the reference value as a
percentage of the reference value
*1
(percent tolerance mode).
*1.Use the :CALC:COMP:PRIM:NOM command to set the reference value.