188 Chapter 10
Command Reference
:SOUR:VOLT
:SOUR:VOLT
Syntax :SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] <numeric>[MV|M|V]
:SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Description Sets the measurement signal level.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that a fraction below the resolution is rounded off.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Level]
<numeric>
Description Measurement signal level
Range 100E-3 to 1
Initial value 1
Unit V
Resolution 0.1