156 Chapter 10
Command Reference
:CORR:CKIT:STAN3:FORM
:CORR:CKIT:STAN3:FORM
Syntax [:SENSe]:CORRection:CKIT:STANdard3:FORMat {CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}
[:SENSe]:CORRection:CKIT:STANdard3:FORMat?
Description Sets the types of parameters used to define the standard for LOAD compensation.
Parameters
For details on CP and CS, refer to :CALC1:FORM on page 147. For details on D, Q, RP,
and RS, refer to :CALC2:FORM on page 150.
Query response {CPD|CPQ|CPG|CPRP|CSD|CSQ|CSRS}<newline><^END>
Related commands :CORR:CKIT:STAN3 on page 155
Equivalent key
sequence
[Load]( [Blue] - [6] ) - CorVal - PrmSlct
Description
CPD
(initial value)
Specifies CP as the primary parameter, D as the secondary parameter.
CPQ Specifies CP as the primary parameter, Q as the secondary parameter.
CPG Specifies CP as the primary parameter, G as the secondary parameter.
CPRP Specifies CP as the primary parameter, RP as the secondary parameter.
CSD Specifies CS as the primary parameter, D as the secondary parameter.
CSQ Specifies CS as the primary parameter, Q as the secondary parameter.
CSRS Specifies CS as the primary parameter, RS as the secondary parameter.