Chapter 10 145
Command Reference
:CALC:COMP:SEC:LIM
10. Command Reference
:CALC:COMP:SEC:LIM
Syntax :CALCulate:COMParator:SECondary:LIMit <numeric 1>,<numeric 2>
:CALCulate:COMParator:SECondary:LIMit?
Description Sets the limit range for the secondary parameter used in the comparator function.
This command only sets the limit range. To activate the set limit range, use the
:CALC:COMP:SEC:STAT command to enable sorting judgment for the measurement
result of the secondary parameter.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands :CALC:COMP:SEC:STAT on page 146
:CALC:COMP:PRIM:BIN{1-9} on page 142
:CALC:COMP on page 134
Equivalent key
sequence
[Sec Low]( [Blue] - [Pri Low]) and [Sec High]( [Blue] - [Pri High] )
[Comprtr]( [Blue] - [1] ) - Limit - Sec
<numeric 1> <numeric 2>
Description The lower limit value of the
limit range.
The upper limit value of the
limit range.
Range -99.9999E9 to 99.9999E9 -99.9999E9 to 99.9999E9
Initial value 0 0
Unit Depends on the type of
parameter.
Depends on the type of
parameter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals