Chapter 10 131
Command Reference
:AVER
10. Command Reference
:AVER
Syntax [:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description Turns ON/OFF the averaging function.
Parameters
Query response {1|0}<newline><^END>
Related commands :AVER:COUN on page 132
Equivalent key
sequence
[Average]( [Blue] - [Meas Time] )
NOTE You cannot turn ON/OFF the averaging function directly by using the front panel keys.
Setting the averaging count with the front panel keys will automatically turn on the
function. Note that after the function has started this way you cannot turn it off by using the
front panel keys.
Description
ON or 1 (initial value) Turns ON the averaging function.
OFF or 0 Turns OFF the averaging function.