134 Chapter 10
Command Reference
:CALC:COMP
:CALC:COMP
Syntax :CALCulate:COMParator[:STATe] {ON|OFF|1|0}
:CALCulate:COMParator[:STATe]?
Description Turns ON/OFF the comparator function.
This setup is interlocked with the ON/OFF state of the signal output of the handler
interface.
NOTE Changing the measurement parameter will automatically turn off this function. To avoid
this, you have to write a program so that this command is executed after the execution of
the measurement parameter setup command (:CALC1:FORM and :CALC2:FORM).
Parameters
Query response {1|0}<newline><^END>
Related commands :CALC1:FORM on page 147
:CALC2:FORM on page 150
Equivalent key
sequence
[Comprtr]( [Blue] - [1] ) - On/Off
Description
ON or 1 Turns ON the comparator function.
OFF or 0 (initial value) Turns OFF the comparator function.