Chapter 10 173
Command Reference
:DATA:FEED
10. Command Reference
:DATA:FEED
Syntax :DATA:FEED {BUF1|BUF2},{"CALCulate1"|"CALCulate2"|""}
:DATA:FEED? {BUF1|BUF2}
Description Selects the measurement data fed into data buffer 1 or data buffer 2 from the following:
primary parameter, secondary parameter, or none to be fed. The query response is a string
with double quotation marks (“).
Parameters
Query response {"CALCulate1"|"CALCulate2"|""}<newline><^END>
Related commands :DATA:FEED:CONT on page 174
:DATA:POIN on page 175
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
BUF1 Specifies data buffer 1, to which the
{CALCulate1”|”CALCulate2”|””} setting is applied or reads
out the setup of data buffer 1.
BUF2 Specifies data buffer 2, to which the
{“CALCulate1”|”CALCulate2“|””} setting is applied. Or reads
out the setup of data buffer 2.
Description
“CALCulate1” Specifies the primary parameter as the measurement data fed into
the data buffer specified with {BUF1|BUF2}.
“CALCulate2” Specifies the secondary parameter as the measurement data fed
into the data buffer specified with {BUF1|BUF2}.
““ (initial value) Does not feed the measurement data into the data buffer specified
with {BUF1|BUF2}.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals