140 Chapter 10
Command Reference
:CALC:COMP:COUN:MULT:OVLD?
:CALC:COMP:COUN:MULT:OVLD?
Syntax :CALCulate:COMParator:COUNt:MULTi:OVLD?
Description Reads out the overload count value of the selected channel.
When the multi-compensation function is tuned off, reads out normal count value
(common value for all channels). In other words, this command has the same function as
the :CALC:COMP:COUN:OVLD? command. (Query only)
Query response {numeric}<newline><^END>
Related commands :CALC:COMP:COUN:MULT:DATA? on page 139
:CORR:MULT on page 161
:CALC:COMP:COUN on page 137
:CALC:COMP:COUN:CLE on page 137
:CALC:COMP on page 134
Equivalent key
sequence
No equivalent key is available on the front panel.
:CALC:COMP:COUN:OVLD?
Syntax :CALCulate:COMParator:COUNt:OVLD?
Description Reads out each overload count value. (Query only)
Query response {numeric}<newline><^END>
Related commands :CALC:COMP:COUN:DATA? on page 138
:CALC:COMP:COUN on page 137
:CALC:COMP:COUN:CLE on page 137
:CALC:COMP on page 134
Equivalent key
sequence
No equivalent key is available on the front panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals