Chapter 7 85
Sorting Based on Measured Result (Comparator Function)
Rejecting Excessively Low Measured Results (Low C reject function)
7. Sorting Based on
Measured Result
(Comparator Function)
Rejecting Excessively Low Measured Results (Low C reject
function)
Turning ON/OFF Low C reject function
To turn ON/OFF the Low C reject function, use the following command.
• :CREJ on page 166
You can detect an excessively low (equal to or less than the preset boundary value)
measured result of the primary parameter value (Cp or Cs) as Low C (abnormal
measurement status) by turning ON the Low C reject function.
NOTE When the comparator function is ON, the sorting judgment is performed normally even if
Low C is detected. However, the sorting judgment result displayed on the screen is LOWC
and, on the handler interface, the /LOW_C_REJECT signal becomes active (LOW) in
addition to the sorting judgment signal.
Setting limit (range boundary values) of Low C reject function
To set the limit (boundary values of the range in which Low C is detected) of the Low C
reject function, use the following command.
• :CREJ:LIM on page 167

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals