238 Appendix B
Information for Replacing 4278A with 4288A
At-a-Glance List of GPIB Command Correspondence
At-a-Glance List of GPIB Command Correspondence
Table B-25 (alphabetical order of 4278A commands) and Table B-26 (by function of
4278A commands) provide at-a-glance lists of the GPIB command correspondence
between the 4278A and the 4288A. For details on the command correspondence, refer to
“Functional Comparison and GPIB Command Correspondence” on page 212.
Table B-25 At-a-glance list of GPIB command correspondence between 4278A and 4288A
(alphabetical order)
4278A Function overview 4288A
[ A ]
AOFF{0|1} Turns ON/OFF the offset compensation for the primary parameter. :CORR:OFFS on page 163
AUX{0|1} Turns ON/OFF the AUX BIN function. :CALC:COMP:AUXB on page 135
AVE= Sets the averaging count. :AVER:COUN on page 132
[ B ]
BIN{1-9}= Sets the limit range of each BIN in the tolerance mode. :CALC:COMP:PRIM:BIN{1-9} on page 142
BLIM= Sets the limit range in the sequential mode.
BOFF{0|1} Turns ON/OFF the offset compensation for the secondary parameter. :CORR:OFFS on page 163
[ C ]
CABL{0-2} Sets the cable length. :CAL:CABL on page 133
CCOU? Reads out the BIN count value of each channel. :CALC:COMP:COUN:MULT:DATA? on
page 139
:CALC:COMP:COUN:MULT:OVLD? on
page 140
CLIM Clears all limit ranges. :CALC:COMP:CLE on page 136
CNO= Sets the channels for multi-compensation. :CORR:MULT:CHAN on page 161
CNT{0|1} Turns ON/OFF the BIN count function. :CALC:COMP:COUN on page 137
COMP{0|1} Turns ON/OFF the comparator function. :CALC:COMP on page 134
COUN? Reads out the BIN count value. :CALC:COMP:COUN:DATA? on page 138
:CALC:COMP:COUN:OVLD? on page 140
CREJ0 Turns OFF the Low C reject function. :CREJ on page 166
CREJ{1-6} Sets the Low C detection boundary value for the Low C reject function. :CREJ:LIM on page 167
CSTD= Sets the Cp value of the standard for the standard (LOAD) compensation. :CORR:CKIT:STAN3 on page 155
[ D ]
DATA? Reads out the measurement result. :FETC? on page 181
DBIN{0|1} Turns ON/OFF the simultaneous output of the comparator sorting result
when reading out the measurement result.
N/A (dependent on the ON/OFF of the
comparator function)
DDIG{4-6} Sets the number of display digits for the measurement result. :DISP:TEXT1:DIG on page 176
DEND Turns OFF the continuous data output (spew-out mode). N/A
DFMT{1|2} Sets the data output (data transfer) format. :FORM on page 182
DPAG{1-4} Sets the display format on the display. N/A
DST Turns ON the continuous data output (spew-out mode). N/A