162 Chapter 10
Command Reference
:CORR:MULT:CKIT:STAN3
:CORR:MULT:CKIT:STAN3
Syntax [:SENSe]:CORRection:MULTiple:CKIT:STANdard3[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:MULTiple:CKIT:STANdard3[:STATe]?
Description Determines whether to define the standard values for LOAD compensation for each
channel individually when using the multi-compensation function (set to ON with the
:CORR:MULT command).
Parameters
Query response {1|0}<newline><^END>
Related commands :CORR:MULT on page 161
:CORR:CKIT:STAN3 on page 155
Equivalent key
sequence
[Scanner]( [Blue] - [9] ) - LoadRef
Description
ON or 1 Enables channel-by-channel value definition.
OFF or 0
(initial value)
Disables channel-by-channel value definition (defines the same
values for all channels).