Chapter 10 127
Command Reference
*SAV
10. Command Reference
*SAV
Syntax *SAV <numeric>
Description Saves the instrument setups to the register of the specified number on EEPROM. For
details of saved instrument setups, refer to Appendix D, “Initial Settings.” (No query)
Parameters
If the specified parameter is out of the allowable setup range, an error occurs.
Related commands *RCL on page 126
Equivalent key
sequence
[Save]( [Blue] - [Rcl] )
*SRE
Syntax *SRE <numeric>
*SRE?
Description Sets the value of the Service Request Enable Register.
Parameters
If the specified parameter is out of the allowable setup range, the result of bitwise AND
with 255 (0xff) is set. Note that bit 6 cannot be set to 1.
Query response {numeric}<newline><^END>
Related commands *ESE on page 124
:STAT:OPER:ENAB on page 189
Equivalent key
sequence
No equivalent key is available on the front panel.
<numeric>
Description Specified number
Range 0 to 9
Resolution 1
<numeric>
Description Setup value of the register
Range 0 to 255
Initial value 0
Resolution 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals