Appendix B 241
Information for Replacing 4278A with 4288A
At-a-Glance List of GPIB Command Correspondence
B. Information for Replacing
4278A with 4288A
Table B-26 At-a-glance list of GPIB command correspondence between 4278A and 4288A
(by function)
Item to Be Set Up/Executed 4278A 4288A
Measurement
condition
Reset *RST :SYST:PRES on page 193
Measurement parameter setup MPAR{1-6} :CALC1:FORM on page 147,
:CALC2:FORM on page 150
Measurement signal
setup
Measurement frequency FREQ{1|2} :SOUR:FREQ on page 187
OSC level (Measurement signal
level)
OSC= :SOUR:VOLT on page 188
HI-ACC mode setup ON/OFF HIAC{0|1} N/A (The HI-ACC mode is not available.)
Measurement range value RC=
Measurement range
setup
When the measurement frequency
is 1 kHz:
RA{0-7} :RANG on page 184,
:RANG:AUTO on page 185
When the measurement frequency
is 1 MHz:
RB{0-11}
Integration time (measurement time) setup ITIM{1-3} :APER on page 130
Averaging count setup AVE= :AVER:COUN on page 132
Cable length setup CABL{0-2} :CAL:CABL on page 133
Delay time (trigger delay time) setup DTIM= :TRIG:DEL on page 195
Trigger mode setup TRIG{1-3} :TRIG:SOUR on page 196
Compensation OPEN compensation ON/OFF OPEN{0|1} :CORR:OPEN on page 165
SHORT compensation ON/OFF SHOR{0|1} :CORR:SHOR on page 165
STANDARD (LOAD) compensation ON/OFF STD{0|1} :CORR:LOAD on page 160
Offset compensation
ON/OFF
Primary parameter AOFF{0|1} :CORR:OFFS on page 163
(The primary and secondary parameters
cannot be turned ON/OFF separately.)
Secondary parameter BOFF{0|1}
Standard for the
standard (LOAD)
compensation
Definition parameter setup SPAR{1|2} :CORR:CKIT:STAN3:FORM on page 156
Cp value setup CSTD= :CORR:CKIT:STAN3 on page 155
D value setup DSTD=
G value setup GSTD=
Setup value readout STR?
Measurement to
obtain the measured
compensation value
(data for
compensation)
OPEN XOP :CORR:COLL on page 157
SHORT XSH
Standard (LOAD) XSTD
Offset compensation
value setup
Primary parameter OFFA= :CORR:OFFS:DATA on page 164
Secondary parameter OFFB=
Measured
compensation value
(data for
compensation)
readout
OPEN OPM? :CORR:DATA on page 158
SHORT SHM?
Standard (LOAD) STM?
Temperature compensation execution XTMP N/A (No need for the temperature
compensation.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals