Detectors
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 162
3. Disconnect the heater/sensor leads from the
connector on the right side of the GC.
Figure 1-8. Disconnecting the µECD cables
4. Inside the oven, remove the insulation cup
and disconnect the column from the makeup
gas adapter.
5. Use a 9/16-inch wrench to loosen the
1/4-inch Swagelok nut on the makeup gas
adapter from the bottom of the detector.
Slide the makeup gas adapter out of the
bottom of the detector.
Heater/sensor
Ribbon cable
Need New
cable
connector