EasyManuals Logo

Agilent Technologies 6850 Series User Manual

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
Detectors
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 162
3. Disconnect the heater/sensor leads from the
connector on the right side of the GC.
Figure 1-8. Disconnecting the µECD cables
4. Inside the oven, remove the insulation cup
and disconnect the column from the makeup
gas adapter.
5. Use a 9/16-inch wrench to loosen the
1/4-inch Swagelok nut on the makeup gas
adapter from the bottom of the detector.
Slide the makeup gas adapter out of the
bottom of the detector.
Heater/sensor
Ribbon cable
Need New
cable
connector

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals