Released: April 2004 6850 Series Gas Chromatograph Service Procedures 182
Figure 1-20.Removing the FPD housing from
the support bracket.
7. Use two wrenches to loosen the Swagelok
connection between the jet assembly and the
heated transfer tube.
8. Carefully lift the detector module and jet
assembly off of the transfer tube and out of
the GC. Be careful not to damage the fused
silica gigabore liner.
Vent line
Ignitor
Loosen screws
Ignitor cable
Heater transfer line nut
Heater block
U-clamp screws
(one screw hidden)
Jet assembly
new?